Analysis of Azithromycin with HPLC Polymer Column According to USP: Shodex

10 August 2022

Azithromycin is a 15-membered ring macrolide antibiotic. According to the United States Pharmacopeia and the National Formulary (USP 39-NF 34**), analysis of azithromycin should be carried out with a column filled with L67 packing material and meets following requirements. The Shodex Asahipak ODP-50 4E confirmed the requirements were met.

System suitability requirements:
 Resolution of azithromycin and azaerythromycin A: ≥ 3.0
 Relative standard deviation (RSD): ≤ 1.1%
 Tailing factor of azithromycin: 0.8 to 1.5

**The version at the time of the application acquisition.


Sample : 10 μL; Azithromycin, Azaerythromycin A, dissolved in 6.7 g/L Dibasic potassium phosphate aq. (pH8.0 adjusted with phosphoric acid)/CH3CN=40/60

Column: Shodex Asahipak ODP-50 4E (4.6 mm I.D. x 250 mm)
Eluent: 6.7 g/L Dibasic potassium phosphate aq. (pH11.0 adjusted with 10 M KOH) /CH3CN=40/60
Flow rate: 1.0 mL/min
Detector: UV (210 nm)
Column temp.: 40 °C

Introduction to Reversed-phase Chromatrography

Source: www.shodex.com – Analysis of Azithromycin